Laboratory Products & Solutions

Cyclosizer Quick Calibration Tool

If your Cyclosizer’s C1 – C5 interlinking pipework has been dismantled or if your unit is aging then it will probably need to be re-calibrated to determine its new separating capabilities. 

The calculator below assumes the use of MARC-Supplied calibration samples (CZ229) and also assumes you have read the operation manual and have experience operating the Cyclosizer.  

How to perform a Cyclosizer Calibration:

  1. Weigh a calibration silica sample including bag (CZ229).
  2. Empty the silica into the sample container.
  3. Weigh the empty bag and subtract the mass to calculate the mass of silica in the sample container. 
  4. Add Water to the sample container and continue as per the Operation Manual. 
  5. Note: Perform all calibration runs on the same day to keep the water temperature as similar as possible.  
  6. As per the Operation Manual perform runs at 11.6 L/Min and record the water temperatures of each run.
    • Perform 2 runs at 15, 30 and 45 mins (6 runs in total). 
    • Record the masses of each C1 – C5 fraction after filtering & drying. 
    • If you only require a quick and indicative assessment of your unit’s performance perform 1 run at 15, 30 and 45 mins (3 runs in total). Duplicate the results for each time (i.e. 15, 30 and 45 mins) in the table below so that all cells are populated.
  7. Enter your results into the table below and click “Calculate”. “Load Sample Data” loads a pre-determined arbitrary set of results.
15 min 30 min 45 min
Run # 1 2 3 4 5 6
Net Silica Sample Mass (g)
Water Temp (deg C)

Note: Flow rate for run needs to be set at 11.6 L/Min

C1 Net Sample Mass after Drying (g)
C2 Net Sample Mass after Drying (g)
C3 Net Sample Mass after Drying (g)
C4 Net Sample Mass after Drying (g)
C5 Net Sample Mass after Drying (g)

52 total views, 1 views today